Operation
33
Stat
Pressing the Stat button will:
Enter Statistics mode—if Statistics mode is not
active and the button is pressed for less than two
seconds.
Exit Statistics mode.
Enter Edit mode and select the statistics parameters
to be edited (Batch Size and Restart mode) after
edit cycle initiation (if Setup/Local is pressed
beforehand). Batch size is 2 to 99,999 pulses (ther-
mopile sensors in Joules mode, or pyroelectric
sensors), or 1 to 99,999 seconds (thermopile
sensors in Watts mode, or optical sensors).
View the statistics parameters if the button is
pressed for two seconds or more.
Wave
You can configure FieldMaxII-TOP to automati-
cally account for any difference between the laser
wavelength and the calibration wavelength. In the
case of optical sensors, this compensation is necessary
because the sensor contains calibration data from a
number of different wavelengths. Thermopile and
optical sensors include wavelength compensation infor-
mation that is used in this mode.
After pressing Setup/Local, the Wave button is used to
enter Edit mode and set the wavelength. If an edit cycle
has not been initiated, pressing the Wave button will