Host Interface
69
propget SpeedupAnalogOutput([out, retval]
VARIANT_BOOL* );
The read only property indicating mode of application of
thermopile speedup to the analog output channel
(enabled or disabled).
propget SpeedupDigitalDisplay([out, retval]
VARIANT_BOOL* );
The read only property indicating mode of application of
thermopile speedup to the digital display (enabled or
disabled).
propget SpeedupHostData([out, retval]
VARIANT_BOOL* );
The read only property indicating mode of application of
thermopile speedup to the host data output channel
(enabled or disabled).
propget SpeedupMeter([out, retval] VARIANT_BOOL*
);
The read only property indicating mode of application of
thermopile speedup to the simulated analog meter
(enabled or disabled).
propget StatisticsMode([out, retval] short* );
The read only property indicating the statistics mode.
propget TriggerLevel([out, retval] short* );
The read only property indicating pulse trigger level.
This property is meaningful only when a pyroelectric
probe is attached.
propget Wavelength([out, retval] long* );
The read only property indicating the wavelength of
operation.